Recall Z-0811-2010 ยท MKJ Automated External Defibrillators (Non-Wearable)

Physio Control, Inc.

Biphasic LIFEPAK 12 defibrillator / monitor.

DateFeb 22, 2010 (initiated Mar 31, 2009)
StatusTerminated
ReasonAffected devices contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock. There could be a delay in therapy or prevention of defibrillation therapy if the defibrillator were needed in a cardiac arrest situation. Delay or prevention of defibrillation therapy may contribute to the death of a viable patient.
Root causeNonconforming Material/Component
ActionConsignees were sent a notification letter dated March 2009 that explained firm representatives would follow up with each customer within 30 days to arrange an update to affected devices. Affected Biphasic PCBAs will be replaced. Customers advised to keep the devices in use and to continue to perform daily User Test.
Quantity1718
DistributionDistributed domestically and internationally to hospitals, clinics, etc. Foreign distribution to the following countries: Australia, Brazil, Canada, Colombia, Costa Rica, Germany, Hong Kong, Mexico, Netherlands, and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063510LIFEPAK 12 DEFIBRILLATOR/MONITOR
Medtronic Emergency Response Systems, Inc., Jan 26, 2007