Recall Z-1271-2010 ยท LKD Monitor, Carbon-Dioxide, Cutaneous

Philips Healthcare Inc.

French 100-120V 860 TC C02 & 02 Monitor, Respironics/Philips product. Used for trending O2 and CO2 tension at the skin surface for neonates and adults as an adjunct to arterial O2 and CO2 measurements.

DateApr 8, 2010 (initiated Nov 19, 2009)
StatusTerminated
ReasonPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may crack and fail at/or inside the plug. Potential for burns or electrical shock.
Root causeNonconforming Material/Component
ActionPhilips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may crack and fail at/or inside plug. The issue affects medical device power cords equipped with a plug that has a prong and ground-pin insert design and a black plastic bridge connecting the terminal prongs on the plug ("taller bridge" design). Medical device power cords that do not have a black bridge connecting the terminal prongs on the plug are not affected. Urgent Medical Device Correction let
Quantity0
DistributionWorldwide Distribution -- UNITED STATES, ALGERIA, ARGENTINA, Aruba, AUSTRALIA, BAHRAIN, BANGLADESH, BELGIUM, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, Denmark, ECUADOR, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, HUNGARY, INDIA, INDONES
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K890517TRANSCUTANEOUS O2/CO2 MONITOR
Novametrix Medical Systems, Inc., Apr 14, 1989