Recall Z-0546-2010 ยท HBB Motor, Drill, Pneumatic

Medtronic Sofamor Danek Usa, Inc - Dallas Distribution

Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotomy applications.

DateDec 17, 2009 (initiated Nov 30, 2009)
StatusTerminated
ReasonComponent of product attachment may be outside established tolerance with potential for footed portion to break during normal use.
Root causeNonconforming Material/Component
ActionMedtronic issued an "Urgent: Medical Device Recall" letter dated November 30, 2009. Consignees were notified of the issued and asked to quarantine any affected product and return affected product to the firm for replacement. For further information, contact Medtronic Powered Surgical Solutions Customer Service at 1-800-433-7080 or 1-817-788-6400.
Quantity64
DistributionWorldwide Distribution -- Including the United States (DC, TN, NC, KY, AL, FL, NJ, OH, MI, IN, NY, WA, CA, NV, NM, IA, MO, NE and UT), Great Britain, Brazil, Turkey, Singapore and the Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020069MEDTRONIC MIDA REX LEGEND SYSTEM, MIDAS REX LEGEND PNEUMATIC HIGH SPEED SYSTEM OR LEGEND SYSTEM
Medtronic Midas Rex, Mar 18, 2002