ClearanceBrief
Recall Z-1238-2010 ยท FQO Table, Operating-Room, Ac-Powered

Stryker Communications Corp

Vertier Surgical Table, Catalog numbers 0788100000, 0788100000S, 0788100001, and 0788100001S

DateApr 6, 2010 (initiated Nov 13, 2009)
StatusTerminated
ReasonIn certain positions, product is susceptible to fluid instrusion which could possibly result in short circuit or unsolicited movement.
Root causeOther
ActionFirm notified all consignees by letter sent 11/13/2009. Contact the firm at 972-410-7100 with questions relating to this recall.
Quantity206
DistributionUnited States (GA, AR, ND, KY, CO, CA, SD, PA, OH, TX, SC, NY, MS, IL, KS, NJ, OR, WA, MN, VA, IL, OK, WY, IN, ID, UT, MO, FL, and DC) and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.