Recall Z-1015-2013 ยท BRY Cabinet, Table And Tray, Anesthesia

CareFusion Corporation

Pyxis Anesthesia System Model 2000 (PAS 2000) Product Usage: The PAS 2000 is an anesthesia medication dispensing device that provides medications and inventory management for the anesthesiologist at the point of care within the operating room.

DateMar 28, 2013 (initiated Nov 20, 2009)
StatusTerminated
ReasonThe recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up.
Root causeDevice Design
ActionCarefusion sent a Safety Alert Notification letter dated January 13, 2010 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to ensure back-up procedures were in place in case of system down-time or lock-up, including review of the back-up procedures with staff. The letter included an update to the PAS 2000 User Guide. The letterstates that a CareFusion field representative will be contacting customers to make arrangements to apply warning labels to the PAS 2000 electronics doors. Customers were instructed to c
Quantity2637
DistributionWorldwide Distribution - USA (Nationwide) including CANADA.
FDA recordOpen on fda.gov