Recall Z-0567-2010 ยท CAW Generator, Oxygen, Portable

Sequal Technologies Inc

SeQual Technologies Inc. Eclipse Oxygen System, Model 1000 Indicated for the administration of supplemental oxygen.

DateDec 23, 2009 (initiated Oct 1, 2009)
StatusTerminated
ReasonUnit may shut down when operating from the Power Cartridge. A temporary power interruption would cause the unit to sound an alarm and shut down. If the patient does not have access to an alternate supply of oxygen or access to AC or DC power, this may lead to shortness of breath, hypoxia, or in some patients may cause serious health consequences.
Root causePending
ActionThe recall notification was initiated on November 12, 2009 with the firm forwarding a "Medical Device Correction" letter to all of the affected consignees via US Postal Service for US Consignees and FedEx for Foreign Consignees. The Medical Device Correction Letter contained the following: (1) Medical Device Correction Letter; (2) List of Serial Number affected that were distributed to the Provider; (3) Reply Card that the Provider will be requested to send back regarding receipt notification of the Medical Device Correction Letter and confirming that they have the affected product; (4) Addi
Quantity6,588 units
DistributionWorldwide Distribution -- Including countries of USA, ARGENTINA, AUSTRIA, BAHAMAS, BERMUDA, CANADA, CHILE, COLUMBIA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, ITALY, ISRAEL,JAPAN, JORDAN, KUWAIT, MALTA, NETHERLANDS, POLAND, PUERTO RICO, SAUDI ARABIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, TAIWAN,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013931OMNI OXYGEN SYSTEM, MODEL 1000
Sequal Technologies, Inc., Mar 27, 2002