Recall Z-1751-2010 ยท NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2

DiaDexus, Inc

PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 90115, Beckman test kit, manufactured by DiaDexus, South San Francisco, CA.

DateJun 3, 2010 (initiated Oct 29, 2009)
StatusTerminated
ReasonInaccurate results-- some samples may return with falsely elevated results potentially as a result of an interfering substance. Some patients may be subjected to additional lifestyle changes or intensified pharmaceutical therapy; such as statins, as a result of the falsely elevated result.
Root causeOther
ActionNotification letter issued to consignees on 10/29/2009. Customers were instructed to not report PLAC test results for samples that are greater than 360 ng/mL, and pursue confirmatory testing by contacting DiaDexus. Customers with questions please call the company technical support line toll free at 1-877-752-2837 for assistance. UPDATE: On April 30, the firm suspended marketing of the PLAC TIA method, as the limited range no longer covered the clinical data/reference range data cited in the product insert. The firm sent out a letter notifying customers on May 10, 2010. The firm set up a
Quantityall lots
DistributionProduct was distributed to 123 consignees throughout the US.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072599PLAC TEST REAGENT KIT,CALIBRATOR KIT, LP-PLA2 CONTROL KIT, MODEL(S) 90107,90109,90109
Diadexus, Inc., Dec 20, 2007