Recall Z-0589-2010 ยท CBK Ventilator, Continuous, Facility Use

MAQUET Inc.

Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden

DateJan 5, 2010 (initiated Nov 17, 2009)
StatusTerminated
ReasonSome Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
Root causeDevice Design
ActionMaquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
Quantity29364 (8438 of those in US)
DistributionWorldwide distribution