Recall Z-2384-2010 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Whatman, Inc.

Whatman Body Fluid Collection Paper- 903 California Newborn Screening Card Ref 110534 903 Card are intended for use by healthcare professionals in the collection and in-vitro storage of neonate human blood. The neonate blood is tested to screen the infant for congenital abnormalities of metabolism and other conditions.790

DateSep 7, 2010 (initiated Oct 29, 2009)
StatusTerminated
ReasonNeonate newborn screening card demographic and filter portion may be detached
Root causeProcess control
ActionThe firm, Whatman, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 29, 2009, to the customer. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: inspect the card and ensure the 903 Paper Portion is firmly and completely fixed to the Demographic portio, prior to recording patient demographic details on the card and obtaining a patient sample; discard the card, obtain a new card and inspect as above, if the 903 Paper Portion of the card partially or completely detaches from the Demographic portion of the car
Quantity488,000 cards
DistributionNationwide distribution: CA