Recall Z-0488-2010 ยท EKM Gutta-Percha

Sybronendo

Elements Gutta Percha and Elements RealSeal Cartridges; Part Number: 972-1002, 972-1003, 972-2001, 972-2002

DateDec 9, 2009 (initiated Oct 27, 2009)
StatusTerminated
ReasonThe recall was initiated because during the manufacturing process of the affected units, the molding process resulted in lock nuts that did not have the level of acceptability required, and therefore, may not withstand the pressures exhibited by the Elements Obturation Unit with which the cartridges are used. More Specifically, there exists a possibility that if the lock nut were to fail, the cart
Root causeComponent design/selection
ActionThe recall communications was initiated on November 06, 2009 with the firm forwarding a Recall Letter to the affected consignees via US Postal Service 1st class mail. The recall letter informed the consignees of the reason for recall, the hazard and the affected products. Consignees were instructed to call SybronEndo Customer Care at 1-800-346-3636 directly to handle returns & replacements. In addition, consignees were asked to complete the enclosed Recall Return Form and return it by fax to 909-962-5605. The recall notice also instructed consignees that are an authorized SybronEndo distribut
Quantity5046 units (3720 units in US)
DistributionNationwide and Worldwide to: USA, Canada, Chile, China, Ecuador, Guatemala, Hong Kong, Japan, Malaysia, Singapore, Taiwan and Venezuela
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023819FIBERFILL SGP
Pentron Clinical Technologies, Mar 26, 2003