Recall Z-0662-2010 ยท IOQ Bed, Flotation Therapy, Powered

KCI USA, Inc.

AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord #4100079 used with Kinair IV Model 22600, Kinair IV 60601 Model 216050, Triadyne Proventa Model 406000, Therapulse ATP/II Models 227800 and 227080.S, Kinair Medsurg Model 201001W, and Kinair Medsurg Pulse Model 40900.

DateJan 25, 2010 (initiated Nov 4, 2009)
StatusTerminated
ReasonPower cords may crack and fail inside plug with potential for fire hazard.
Root causeNonconforming Material/Component
ActionKinetics Concepts, Inc. (KCI) issued an "Urgent Medical Device Correction" letter dated November 4, 2009. Consignees were informed of the affected product and instructed to take necessary action to avoid any further issues associated with the product. For further information, contact KCI's technical support team at 1-800-275-4524 option #3.
Quantity4,637 units
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.