Recall Z-1007-2011 ยท ITX Transducer, Ultrasonic, Diagnostic
Siemens Medical Solutions USA, Inc.
10430054 Acuson with Physio Module, manufactured by Siemens Medical Solutions, Mountain View, CA.
| Date | Feb 1, 2011 (initiated Oct 2, 2009) |
| Status | Terminated |
| Reason | ECG waveform does not accurately align with 2D, M-Mode or Spectral Images. |
| Root cause | Nonconforming Material/Component |
| Action | The firm's Customer Safety Advisory notice was mailed to all affected customers. The letter describes when the malfunction occurs and the potential risk. It also recommended that users collect corresponding evidence with spectral waveforms that would confirm or deny findings based on ECG, in order to avoid potential risks for misdiagnosis. The firm planned to exchange the faulty hardware component in the system. Customers would be contacted to schedule for the hardware exchange. Questions regarding this issue should be directed to the local service support person. |
| Quantity | 67 units total |
| Distribution | Worldwide Distribution |