Recall Z-1006-2011 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic
Siemens Medical Solutions USA, Inc.
10041461 ACUSON S2000, manufactured by Siemens Medical Solutions, Mountain View, CA. Intended for Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/Gyn, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
| Date | Feb 1, 2011 (initiated Oct 2, 2009) |
| Status | Terminated |
| Reason | ECG waveform does not accurately align with 2D, M-Mode or Spectral Images. |
| Root cause | Nonconforming Material/Component |
| Action | The firm's Customer Safety Advisory notice was mailed to all affected customers. The letter describes when the malfunction occurs and the potential risk. It also recommended that users collect corresponding evidence with spectral waveforms that would confirm or deny findings based on ECG, in order to avoid potential risks for misdiagnosis. The firm planned to exchange the faulty hardware component in the system. Customers would be contacted to schedule for the hardware exchange. Questions regarding this issue should be directed to the local service support person. |
| Quantity | 67 units total |
| Distribution | Worldwide Distribution |