Recall Z-0910-2011 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Oncology Systems

C-Series Clinac, Trilogy and Novalis Tx used for SRS treatments, Model Numbers H14, H18, H27, H29, HCX, manufactured by Varian Medical Systems, Palo Alto, CA. Device is a radiation Therapy accelerator intended to deliver megavoltage x-ray treatments for conventional radiotherapy (three dimensional conformal radiotherapy and intensity modulated radiotherapy) and stereotactic radiosurgery and radi

DateJan 18, 2011 (initiated Oct 8, 2009)
StatusTerminated
ReasonProduct may deliver radiation treatment to areas larger than intended,to healthy tissue.
Root causeNonconforming Material/Component
ActionVarian sent an URGENT MEDICAL DEVICE CORRECTION letter dated October 8, 2009, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Varian recommended to consignees that test runs be performed prior to actual patient treatment to ensure that the entire system is behaving as intended and that appropriate quality assurance procedures should be followed prior tothe delivery of any treatment plan involving radiation therapy. Operators should monitor the patient and the treatment console continuously during treatment. For quest
Quantity5208 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070094TRILOGY RADIOTHERAPY DELIVERY SYSTEM AND TRILOGY TX DELIVERY SYSTEM
Varian Medical Systems, Inc., Feb 9, 2007
K913119VARIAN C-SERIES CLINACS (CLINAC 2300C/D)
Varian Assoc., Inc., Nov 7, 1991