Recall Z-0481-2010 ยท MRZ Accessories, Pump, Infusion

Animas Corporation

Animas ezManager Max Diabetes Management System version 2.0.9

DateDec 3, 2009 (initiated Nov 2, 2009)
StatusTerminated
ReasonPrinted hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect.
Root causeSoftware design
ActionAnimas issued an Urgent Medical Device Correction letter dated 11/2/09 to the pump user and another letter to the health care professional that prescribed the pump. The letters explain the issue. The patient letter is accompanied with a CD that contains the upgraded software and instructions to follow.
Quantity20,648 units
DistributionNationwide distribution, and Canada, Cayman Islands, and Denmark.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080587EZ MANAGER MAX DIABETES MANAGEMENT SOFTWARE
Animas Corp., May 15, 2008