ClearanceBrief
Recall Z-0409-2010 ยท GDW Staple, Implantable

Surgical Devices/Covidien

Intelligent Battery Pack, product code iB100; Intelligent Battery Charger and Power Supply, product code iB1C1; Intelligent Battery Insertion Guide, product code iBIG-1; Intelligent Battery Insertion Tool, product code iBIT-1.

DateNov 23, 2009 (initiated Sep 23, 2009)
StatusTerminated
ReasonLack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in extended surgical time for the patient.
Root causeNonconforming Material/Component
ActionThe recalling firm issued an Urgent Medical Device Recall letter dated 9/22/09 to all customers. The letter informed customers that Covidien acquired PMI, about the problem with the products, and about the need to return any products remaining inventory. Customers were asked to return the entire inventory of Power Medical Interventions product, regardless of its expiration date. This process can be initiated by contacting Power Medical Innovations customer service at 1-866-769-3763.
Quantity3427 units
DistributionThe products were shipped medical facilities nationwide. The product was also shipped to distributors in Canada, Japan, Germany, Ireland, Italy, Spain, Taiwan, Belgium, Sweden, Netherlands, Israel, Portugal, India, Austria, Turkey, and Greece.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091215INTELLIGENT POWER UNIT, INTELLIGENT CONSOLE, INTELLIGENT DELIVERY SYSTEM, INTELLIGENT SURGICAL INSTRUMENT
Power Medical Interventions, Inc., May 18, 2009