Recall Z-1392-2011 ยท ITX Transducer, Ultrasonic, Diagnostic
Siemens Medical Solutions USA, Inc.
V5M TEE Transducer, when used with an ACUSON S2000 ultrasound system that has cardiac options installed and is at software version VA10, VA10A, VA10B, VA10C, VA15, VA15A, VA15B, VA16 or VA16A, manufactured by Siemens Medical Solutions, Mountain View, CA. Multipurpose mobile software controlled diagnostic ultrasound system and on-screen display for thermal and mechanical indices related to bio-effe
| Date | Feb 22, 2011 (initiated Aug 28, 2009) |
| Status | Terminated |
| Reason | If the transducer is not disconnected from the system after an automatic shutdown due to thermal limits being exceeded, then the next time it is used, the thermal safety guards do not take effect. Normally, the thermal limit before automatic image freeze (and therefore power shut off) occurs at 43 degrees Centigrade. |
| Root cause | Software design |
| Action | Siemens sent a Customer Safety Advisory Notification letter on August 28, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that their local Customer Service Engineer would install a software update as soon as it becomes available. Customers were instructed to pass the Customer Safety Advisory Notification letter to all who need to be aware of these issues until the corrective action is completed. For questions regarding this recall call 650-969-9112 or 800-422-8766. |
| Quantity | 97 units |
| Distribution | Worldwide Distribution - USA (nationwide) and internationally to Canada, China, Czech Republic, Denmark, Germany, Great Britian, Hong Kong, Indonesia, Italy, Japan, Norway, Poland, Portugal, Russian Federation, Singapore, South Korea, Taiwan, and Turkmenistan. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K082142 | ACUSON S2000 DIAGNOSTIC ULTRASOUND SYSTEM Siemens Medical Solutions USA, Inc., Ultrasound DI, Nov 13, 2008 |