ClearanceBrief
Recall Z-1038-2010 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc.

Dimension Vista V-LYTE Fluids: Standard B/Salt Bridge catalog number K825

DateMar 10, 2010 (initiated Oct 22, 2009)
StatusTerminated
ReasonResults can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor.
Root causeOther
ActionThe recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
Quantity150 units
DistributionThe product was shipped to medical facilities nationwide. The product was also shipped to Australia, Canada, Belgium, France, Germany, Italy, Japan, Malaysia, New Zealand, Netherlands, Portugal, Spain, South Korea, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051087DIMENSION VISTA INTEGRATED SYSTEM
Dade Behring, Inc., Jul 11, 2005