Recall Z-1683-2010 ยท MKJ Automated External Defibrillators (Non-Wearable)

Cardiac Science Corporation

Responder 2000 Defibrillator/Monitor Models: 2025653, 2026109, and 2026114. Intended use: The Responder 2000 is indicated for the termination of certain fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and in

DateMay 26, 2010 (initiated Oct 15, 2009)
StatusTerminated
ReasonThe Operators Manual may not have adequate information to clearly explain some device behaviors. Pressing and holding down the Charge or Shock button on the Responder 2000 for more than 5 seconds causes a "Button Stuck" error. With the "Button Stuck" error, the device will automatically exit Manual mode and enter Monitor mode. The user cannot apply therapy while the device is in Monitor mode.
Root causeOther
ActionThe firm, Cardiac Science, sent an "Urgent-Medical Device Corrective Action" letter dated October 15, 2009, was sent to customers. The letter describes the product, problem and action to be taken by the customers. Cardiac Science requested GE to provide the end users with the Operator's Manual addendum. The customers are instructed to place the Operator's Manual addendum with their Responder 2000 Operator's Manual for future reference. If you have any questions about your Responder 2000, please contact Cardiac Science technical support at +44.161.926.0011 or our EU Representative, Ludger Mo
Quantity1838 devices
DistributionWorldwide distribution: USA and Australia, Brazil, Columbia, Costa Rica, Denmark, Finland, France, Germany, Italy, Korea, Mexico, New Zealand, Panama, Paraguay, Peru, Singapore, South Africa, Spain, and Uruguay.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060934MODIFICATION TO POWERHEART ECD
Cardiac Science Corporation, May 15, 2006