ClearanceBrief
Recall Z-0275-2010 ยท FCG Biopsy Needle

Bard Peripheral Vascular Inc

Bard Magnum Disposable Biopsy Core Needle 16 gauge, Catalog Number: MN1616. Bard Magnum Disposable Biopsy Core Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue tumors.

DateNov 17, 2009 (initiated Oct 8, 2009)
StatusTerminated
ReasonThe recall was initiated after the firm determined that some Magnum Disposable Biopsy Core Needle 16 gauge lots may exhibit an exposed sample chamber, due to extended stylet length resulting from a loose hub connection. The exposed sample chamber will typically result in an inability to obtain the biopsy sample. In the majority of reported cases, there was no injury to the patient or user. When us
Root causeProcess design
ActionThe recall notification was initiated by the firm issuing an urgent medical device notification to it's consignees on October 08, 2009 via USPS requesting them to check all inventory locations for the affected products, remove and return affected products to Bard. The consignees were instructed to complete and return a recall effectiveness check form. Bard provided this 800 number 1-800-321-4254 for any questions consignees may have.
Quantity97, 257 units
DistributionNationwide and internationally to Canada Mexico Algeria Andorra Austria Bahrain Belarus Belgium Croatia Cyprus Czech Republic Denmark Finland France French Guiana Germany Gibraltar Greece Hungary Ireland Israel Italy Ivory Coast Kenya Kuwait Latvia Lebanon Luxembourg Malta
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K934370BIP BIOPSY NEEDLE MODIFICATION
Bip USA, Inc., Mar 7, 1994