ClearanceBrief
Recall Z-0449-2010 ยท KWQ Appliance, Fixation, Spinal Intervertebral Body

Stryker Spine

Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.5 x 52MM, spinal implant component

DateNov 24, 2009 (initiated Oct 2, 2009)
StatusTerminated
ReasonSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break at the tip.
Root causeDevice Design
ActionRecall notification letters were sent to all Stryker branches, Hospital Risk Management and Surgeons on October 19, 2009 by Federal Express. Questions are to be directed to Tiffani Rogers, Regulatory Compliance Manager at 201-760-8206.
Quantity165
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073437THOR ANTERIOR PLATING SYSTEM
Stryker Spine, Feb 13, 2008