Recall Z-1146-2010 ยท MVC System, Test, Her-2/Neu, Ihc
Ventana Medical Systems Inc
VENTANA PATHWAY anti-HER-2/neu (4B5) RABBIT MONOCLONAL PRIMARY ANTIBODY (PATHWAY HER2 (4B5)); Model Number: 790-2991, Lot number: 680957, in vitro diagnostic.
| Date | Mar 16, 2010 (initiated Sep 17, 2009) |
| Status | Terminated |
| Reason | Incorrect result: Light staining inconsistencies with lot number 680957 of the Ventana PATHWAY HER-2/neu (4B5) Rabbit Monoclonal Primary Antibody. |
| Root cause | Nonconforming Material/Component |
| Action | The recall was initiated on 09/17/2009 with Ventana Medical Systems forwarding Urgent Voluntary Medical Device Recall letters to all the affected customers advising them to discontinue the use of lot number 680957 and have their laboratory retest with product from a different lot of PATHWAY HER2 (4B5) any breast cases scored as 0, 1+, or 2+ that were not reflexed to an approved in situ hybridization (ISH) method. Customers were also instructed to contact Ventana Customer Support at 1-800-227-2155 and select 1 for Technical Support to speak to a Customer Support agent who will work with then t |
| Quantity | 433 units |
| Distribution | Worldwide distribution: USA, Austria, Denmark, Germany, Hungary, Netherlands, Sweden, United Kingdom, Brazil, Canada, India, Korea, Morocco & Thailand |
| FDA record | Open on fda.gov |