ClearanceBrief
Recall Z-0085-2010 ยท DXC Clamp, Vascular

Edwards Lifesciences Research Medical, Inc.

EndoClamp aortic catheter label reading in part: "CARDIOVATIONS EndoClamp aortic catheter 65 cm,STERILE R, REF EC65, Edwards Lifesciences LLC Irvine, CA 92614-5686" The device is a 10.5Fr wirewound, three-lumen catheter with a balloon near its tip for occluding the ascending aorta. Indicated for use in patients undergoing cardiopulmonary bypass surgery. The catheter occludes the ascending aorta,

DateNov 6, 2009 (initiated Sep 24, 2009)
StatusTerminated
ReasonSpontaneous balloon rupture during use of the product.
Root causeProcess control
ActionCustomers were notified by an Urgent - Product Recall letter, dated 9/29/09, and instructed to check their inventory, identify any unused product and return it to the company. Customers are to contact Customer Service Organization at 1-800-424-3278 to obtain a Returned Goods Authorization number, information concerning replacement product, and if they have any questions.
Quantity2,737 total
DistributionWorldwide Distribution -- US, Europe, Canada, Australia, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955132HEARTPORT ENDOAORTIC CLAMP
Heartport, Inc., Jun 7, 1996
K962510HEARTPORT ENDOAORTIC CLAMP
Heartport, Inc., Sep 11, 1996
K964161HEARTPORT ENDOAORTIC CLAMP
Heartport, Inc., Apr 4, 1997
K974175HEARTPORT ENDOAORTIC CLAMP CATHETER
Heartport, Inc., Dec 16, 1997