Recall Z-0084-2010 ยท DXC Clamp, Vascular
Edwards Lifesciences Research Medical, Inc.
EndoClamp aortic catheter. Label reading in part: "***CARDIOVATIONS EndoClamp aortic catheter 100 cm, REF EC1001, Sterile EO, Sterile R, Edwards Lifesciences LLC Irvine, CA***". The device is a 10.5Fr wirewound, three-lumen catheter with a balloon near its tip for occluding the ascending aorta. Indicated for use in patients undergoing cardiopulmonary bypass surgery. The catheter occludes the as
| Date | Nov 6, 2009 (initiated Sep 24, 2009) |
| Status | Terminated |
| Reason | Balloon catheters were spontaneously rupturing during use. |
| Root cause | Process control |
| Action | Customers were notified by an Urgent - Product Recall letter, dated 9/29/09, and instructed to check their inventory, identify any unused product and return it to the company. Customers are to contact Customer Service Organization at 1-800-424-3278 to obtain a Returned Goods Authorization number, information concerning replacement product, and if they have any questions. |
| Quantity | 2,737 total |
| Distribution | Worldwide Distribution -- US, Europe, Canada, Australia, and Thailand. |