Recall Z-0084-2010 ยท DXC Clamp, Vascular

Edwards Lifesciences Research Medical, Inc.

EndoClamp aortic catheter. Label reading in part: "***CARDIOVATIONS EndoClamp aortic catheter 100 cm, REF EC1001, Sterile EO, Sterile R, Edwards Lifesciences LLC Irvine, CA***". The device is a 10.5Fr wirewound, three-lumen catheter with a balloon near its tip for occluding the ascending aorta. Indicated for use in patients undergoing cardiopulmonary bypass surgery. The catheter occludes the as

DateNov 6, 2009 (initiated Sep 24, 2009)
StatusTerminated
ReasonBalloon catheters were spontaneously rupturing during use.
Root causeProcess control
ActionCustomers were notified by an Urgent - Product Recall letter, dated 9/29/09, and instructed to check their inventory, identify any unused product and return it to the company. Customers are to contact Customer Service Organization at 1-800-424-3278 to obtain a Returned Goods Authorization number, information concerning replacement product, and if they have any questions.
Quantity2,737 total
DistributionWorldwide Distribution -- US, Europe, Canada, Australia, and Thailand.