Recall Z-0544-2010 ยท KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Terumo Cardiovascular Systems Corporation

Sarns Centrifugal Pump (with X-coating), catalog 164275X, packaged in 8 piece shelf boxes, and convenience kits with single pump units. Indicated as extracorporeal blood pumps for use in cardiopulmonary bypass procedures lasting up to 6-hours.

DateDec 17, 2009 (initiated Jul 6, 2009)
StatusTerminated
ReasonThe tubing- to- pump connection in a medical device used in surgical procedures may disengage resulting in blood loss.
Root causeDevice Design
ActionTerumo Cardiovascular Systems Corporation notified consignees of an "Urgent Medical Device Correction" by letter on July 06, 2009. The letter was addressed to hospital Perfusion, Cardiovascular Surgery, and Operating Room departments and contained specific measures to be taken to ensure proper use of the affected device. The firm recommended that users continue to use the devices and follow the corrective measures enclosed with the notification. Consignees were requested to route the notification to all users and confirm receipt of the notification by returning a customer response form. For f
Quantity20,702 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020998SARNS CENTRIFUGAL PUMP WITH XCOATING, MODEL 164275X
Terumo Cardiovascular Systems Corp., Apr 4, 2002