Recall Z-0087-2010 ยท DYB Introducer, Catheter

Cordis Corporation

Cordis CROSSOVER. Cordis CROSSOVER Sheath Introducer. Thomas Medical Products, Inc. 65 Great Valley Parkway, Malvern, PA 19355 USA. Distributed by Cordis Corporation, 14201 North West 60th Avenue, Miami Lakes, Florida 33014 USA Vital Signs, Ltd. 13-14 Eldon Way Lineside Industrial Estate, Littlehampton, West Sussex, UK. The Cordis CROSSOVER Sheath Introducer is intended for use in arterial an

DateNov 5, 2009 (initiated Oct 15, 2009)
StatusTerminated
ReasonCordis has received complaints of stretching or fracture of the sheath shaft during use.
Root causeComponent design/selection
ActionA Medical Device Recall letter dated October 19, 2009 was issued to consignees by sending an overnight letter to each account. The letter described the affected device, issue and actions for customers. A Cordis Corporation representative will follow-up and contact consignees as necessary to facilitate obtaining signature, faxing the acknowledgement form to the Cordis Corporation, collecting and returning units. A press release from the Cordis Corporation announced a nationwide recall of the CROSSOVER Sheath Introducer available on (http://www.webwire.com/ViewPressRel.asp?aId=106460). In
Quantity2,362 units
DistributionWorldwide Distribution -- United States, Belgium, Switzerland, Germany, Finland, France, Italy, Netherlands, Sweden and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083269CROSSOVER
Thomas Medical Products, Inc., Feb 3, 2009