Recall Z-0174-2010 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license

DateNov 10, 2009 (initiated Sep 30, 2009)
StatusTerminated
ReasonIncomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip file on the NAS or HSM will contain fewer images-those which have been archived until the time of the error/interruption.
Root causeSoftware design
ActionThe recalling firm issued a "Customer Safety Advisory Notice" dated 9/14/09 to all affected customers via Update Instructions IM101/09/S. The notice informs the customer of the potential issue and provides instructions to avoid its occurrence. Questions regarding this issue are directed to the local SIEMENS Uptime Service Center.
Quantity2 units
DistributionThe product was shipped to medical facilities in NY and OH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082430SYNGO IMAGING XS
Siemens AG Medical Solutions, Nov 14, 2008