Recall Z-0817-2010 ยท JWX Kit, Screening, Staphylococcus Aureus

Remel, Inc

Oxoid dry Spot Staphytect Plus, Catalog #DR0100M, 120 tests/kit. Oxoid Ltd., Wade road, Basingstoke, Hants, UK. Latex slide agglutination test for the differentiation of staphylococcus aureus by detection of clumping factor, Protein A and certain polysaccarides found in methicillin-resistant S. aureus (MRSA) from those staphylococci that do not possess these properties.

DateFeb 23, 2010 (initiated Jul 20, 2009)
StatusTerminated
ReasonThe products may give false positive results with some non-Staphylococcus aureus strains
Root causeProcess control
ActionThermo Fischer Scientific, Remel Inc. issued letters dated July 20, 2009 via regular mail explaining the reason for recall, risks involved, and requested they discard all remaining product. A Product Inventory Checklist was enclosed to report the amount of product their customer discarded. A second letter dated October 19, 2009 was issued via certified mail. For further information, contact Thermo Fischer Scientific, Remel Inc. at 1-800-255-6730, Extension 3304185.
Quantity27 kits
DistributionNationwide Distribution -- MN, CT, FL, GA, CA, NY, MI, MO, PA, and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.