ClearanceBrief
Recall Z-0102-2010 ยท GEH Unit, Cryosurgical, Accessories

Cooper Surgical, Inc.

WA 4000 Console Model: 900509x.

DateOct 29, 2009 (initiated Sep 29, 2009)
StatusTerminated
ReasonCO2 Cryosystem thermal load may exceed the exhaust gas phase, resulting in ice formation in the exhaust valve. Should this occur, the Cryo tip may cool at a less than optimal rate, resulting in ice formation in the exhaust valve.
Root causeDevice Design
ActionCooper Surgical notified customers by letter titled "Notice of Recall - Urgent" dated September 29, 2009 advising them of the problem and requesting return of the device. International accounts were notified by letter Beginning October 1, 2009. For further information, contact Cooper Surgical by fax at 1-203-799-2002 or phone at 1-203-799-2000.
DistributionWorldwide Distribution -- United States, Canada, Australia, Belgium, England, France, Germany, Greece, Hungary, Iceland, Italy, Ireland, Latvia, Lithuania, Netherlands, Portugal, Spain, Solvenia, Sweden, Hong Kong, Republic of China, Taiwan, Angola, Cameroon, Congo, Croatia, Egypt, Ghana, Guam, Guin
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.