Recall Z-0101-2010 ยท GEH Unit, Cryosurgical, Accessories
Cooper Surgical, Inc.
WA 1000A with Blank Console Model: 900506x.
| Date | Oct 29, 2009 (initiated Sep 29, 2009) |
| Status | Terminated |
| Reason | CO2 Cryosystem thermal load may exceed the exhaust gas phase, resulting in ice formation in the exhaust valve. Should this occur, the Cryo tip may cool at a less than optimal rate, resulting in ice formation in the exhaust valve. |
| Root cause | Device Design |
| Action | Cooper Surgical notified customers by letter titled "Notice of Recall - Urgent" dated September 29, 2009 advising them of the problem and requesting return of the device. International accounts were notified by letter Beginning October 1, 2009. For further information, contact Cooper Surgical by fax at 1-203-799-2002 or phone at 1-203-799-2000. |
| Distribution | Worldwide Distribution -- United States, Canada, Australia, Belgium, England, France, Germany, Greece, Hungary, Iceland, Italy, Ireland, Latvia, Lithuania, Netherlands, Portugal, Spain, Solvenia, Sweden, Hong Kong, Republic of China, Taiwan, Angola, Cameroon, Congo, Croatia, Egypt, Ghana, Guam, Guin |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.