Recall Z-0193-2010 ยท KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Sorin Group USA, Inc.

Revolution Blood Pump with PC Coating, "revOlution Centrifugal pump Catalog# 050300700***Rx ONLY***SORIN GROUP ITALIA via Statale 12 Nord, 86 41037 MIRANDOLA (MO) (Italy) Distributed in the U.S. by: Sorin Group USA, Inc. 14401 W. 65th way Arvada (CO) 80004-3599" Intended for use in cardiopulmonary bypass procedures.

DateNov 13, 2009 (initiated Sep 25, 2009)
StatusTerminated
ReasonA weakness in the plastic packaging tray could cause the tray to crack; compromising product sterility.
Root causePackaging change control
ActionCustomers were notified by an Urgent: Field Safety Notice, dated 9/25/09, and instructed to: check their inventory for the affected product, complete the customer response form and return or visually inspect all affected product before use. For additional information contact Sorin Group Customer Service at 1-800-650-2623.
Quantity256 units
DistributionWorldwide Distribution -- US, Canada, Japan, and Europe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030462COBE REVOLUTION CENTRIFUGAL BLOOD PUMP WITH PC, MODEL 050300700
Cobe Cardiovascular, Inc., Mar 6, 2003