Recall Z-0105-2010 ยท BTR Tube, Tracheal (W/Wo Connector)

Centurion Medical Products

Tri-State Hospital Supply Corp., Centurion Healthcare Products Premie Packs, 50, sterile, Tri-State Hospital Supply Corp., Howell, MI; code no. LM110. Aspiration tray containing Portex uncuffed pediatric sized tracheal tubes.

DateNov 12, 2009 (initiated Sep 22, 2009)
StatusTerminated
ReasonThe pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in an inability to remove secretions and cause partial or complete obstruction of the airway.
Root causeProcess control
ActionThe firm's sales representative was notified of the recall on September 22, 2009 via email and the sole customer was notified by letter dated September 23, 2009. All product was removed from the customer account on September 28, 2009 and returned to the recalling firm.
Quantity650
DistributionIllinois.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.