Recall Z-0263-2010 ยท DZI Drill, Bone, Powered

Stryker Instruments Div. of Stryker Corporation

Stryker Instruments 2.1 mm Cross-Cut Fissure Bur, carbide extended, sterile; REF 2296-101-621.

DateNov 16, 2009 (initiated Oct 5, 2009)
StatusTerminated
ReasonThe cross-cuts are missing on the device, which will result in poor cutting and overheating and/or smoking of the device.
Root causeVendor change control
ActionConsignees were notified by letter dated 10/5/09 and requested to return the devices to Stryker. The international account was notified by email.
Quantity93
DistributionUSA (California, Georgia, Louisiana, Ohio, Wisconsin) and Brazil.