Recall Z-1019-2010 · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Siemens Healthcare Diagnostics Inc

RapidPoint 400 Analyzers for Blood Gas, Electrolytes and metabolites Software Versions prior to 3.5.1. The Siemens 400 series system is intended for point of care and laboratory testing of blood gases, electrolytes and metabolites in arterial, venous and capillary whole blood samples.

DateMar 8, 2010 (initiated Oct 30, 2009)
StatusTerminated
ReasonSoftware: Reference sensor failure may result in inaccurate pH reuslts
Root causeComponent design/selection
ActionAn "Urgent Field Safety Notice" dated September 2009 was sent via overnight mail to customers and included software Version 3.5.2 with instructions for install that will detect and prevent the reporting of potentially errant test results. The customers are requested to please install the included software update immediately and return the attached Fax back form. Siemens will provide the mandatory update to all customers with the Rapidpoint 400 /405 test system. If you have any questions, please contact your local Technical Support at 508-359-3877.
Quantity998 units
DistributionWorldwide distribution: USA and Australia, Austria, Bahamas, Barbados, Belgium, Bosnia Herzeg, Brazil, Canada, Canary Islands, Chile, Croatia, Czech Republic, Egypt, France, Germany, Holy See (Vatican City State) Hong Kong, Ireland, Italy, Japan, Macedonia, Malaysia, M¿xico, New Zealand Not assigned
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002738BAYER RAPIDPOINT 400 SYSTEM
Bayer Corp., Dec 19, 2000