Recall Z-0493-2010 ยท MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Wright Medical Technology Inc

ADVANCE (R) CANAL FILLING STEM EXTENSION, REF: KSP1-2140, DIAMETER 13mm, LENGTH 140mm, 1 EACH, STERILE R, (FOR USE WITH CEMENT ONLY), SINGLE USE ONLY, (Ti6A14V), Wright Medical Technology, Inc., 5677 Airline Road, Arlington, TN 38002

DateDec 8, 2009 (initiated Sep 22, 2009)
StatusTerminated
ReasonTitanium used in the recalled devices was discovered to have internal microstructure conditions of alloy segregation, primary alpha and/or a transformed beta matrix in some of the heat lots. The absence of alloying elements results in an area with less strength. For those instances where the product has been implanted, there is the potential increased risk of failure for these patients.
Root causeProcess control
ActionWright Medical initiated the recall of the product on September 22, 2009 by contacting the distributors by telephone followed with letters addressed to the hospitals and surgeons dated October 9, 2009. At the firm, Customer Service is the contact for return instructions and replacement inventory. This phone number is 800-238-7117. Other questions regarding this matter are directed to 800-874-5630.
Quantity25 units
DistributionNationwide, Columbia, Brazil, Portugal, UK, Belgium, Austria, and Greece
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061223ADVANCE TOTAL KNEE SYSTEM
Wrightmedicaltechnologyinc, Aug 18, 2006