ClearanceBrief
Recall Z-0046-2010 ยท BTM Ventilator, Emergency, Manual (Resuscitator)

Unomedical

Pediatric Manual Pulmonary Resuscitator. Product ID # 12-8708. Packaged under the name of Evergreen.

DateNov 4, 2009 (initiated Oct 6, 2009)
StatusTerminated
ReasonIn some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
Root causeDevice Design
ActionUnomedical issued Press on October 6, 2009. Letters were sent to customers on October 7, 2009 via UPS - return receipt. Letters explained the reason for recall and the potential hazard to patients. The letter contains a list of all affected lot numbers as well as a picture which shows that units with a clearly visible blue ring are NOT affected. Distributors are asked to examine inventory and quarantine all affected product and return the response form, even if no recalled product is in inventory. If the product was further distributed, the letter requests that a copy of the letter be sen
Quantityover 200 total lots
DistributionWorldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K873286DISPOSABLE, MANUAL PULMONARY RESUSCITATOR
Hospitak, Inc., Oct 26, 1987