ClearanceBrief
Recall Z-0185-2010 ยท FTD Lamp, Surgical

MAQUET Inc.

HANAULUX 2004-2005 TV Spring Arm, P/N: 569002998. Intended to provide visible illumination for the surgical field or for the examination of the patient.

DateNov 12, 2009 (initiated Oct 7, 2009)
StatusTerminated
ReasonSeveral reports have been received regarding one model of spring arm (Acrobat 200 series) in combination with surgical light from various Manufacturers. All incidents pertain to a crack on the front joint of the Acrobat arm.
Root causeDevice Design
ActionUrgent Device Correction letters, dated October 7, 2009, were sent out by FedEx. The letter identified the affected products and asked customers to call 1-888-627-8383 #1, if they have any questions. The letter also provided a Technical Description along with a Customer Fax Response form. The letter stated the potential hazard and recommended precaution. If customers need additional info or have any questions they should contact their local MAQUET representative.
Quantity1
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K895715HANAULUX(R) 2000
W.C. Heraeus GmbH, Oct 13, 1989