Recall Z-0509-2010 ยท LIT Catheter, Angioplasty, Peripheral, Transluminal

Numed Inc

IMPACT Catheters, Balloon Dilatation and Esophageal Dilatation Catheter, Model Number 307, STERILE, Rx only --- Common/Usual Name: PTA and Esophageal Catheter --- Manufactured for: B. Braun Interventional Systems Inc., 824 Twelfth Avenue Bethlehem, PA18018 --- Manufactured by: NuMED, Inc. 2880 Main Street, Hopkinton, NY 12965 --- The Impact catheter is packaged in a straight sheath and then

DateDec 15, 2009 (initiated Sep 23, 2009)
StatusTerminated
ReasonThe firm received a complaint from the sole direct account that while performing an inventory review it was observed that the packaging of the IMPACT catheter was cracked and deteriorating. The product was still in its packaging and never used. The customer referred to it as looking like "paint crackle." The risk is that for continued sterility to be assured, the packaging must remain intact.
Root causeLabeling design
ActionAn informal e-mail was sent to B. Braun to give them notice not to distribute any Impact devices until further notice. This was followed up by a formal URGENT MEDICAL DEVICE RECALL letter (dated 9/24/09) to the sole consignee. The products are to be returned to the manufacturer, including any products further distributed. Customer questions should be directed to the recall contact (replace " AT " with @) nlaflesh ATnumedusa.com
Quantity7,367 units domestically
DistributionOne distributor in Pennsylvania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K931009TYSHAK PERIPHERAL BALLOON DILATION CATHETER
NuMED, Inc., Dec 21, 1993
K983373IMPACT BALLOON DILATION CATHETER
B. Braun/Mcgaw, Jun 22, 1999