ClearanceBrief
Recall Z-0004-2010 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Ge Healthcare

GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy.

DateJan 20, 2010 (initiated Aug 12, 2009)
StatusTerminated
ReasonA software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative air kerma to be inaccurate.
Root causeRadiation Control for Health and Safety Act
ActionAn Important Electronic Product Radiation Warning letter will be sent to customers identifying the issues, interim solution, and permanent solution. Questions or concerns are to be directed to the GEHC OEC service team at 1-800-874-7378.
Quantity1411 units total
DistributionNationwide and Canadian Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.