Recall Z-0140-2010 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Kentucky Packaging Service LP

Olsen Medical, MIDAS TOUCH NEEDLE ELECTRODE, WITH MODIFIED TIP. Model Numbers: 30-1011, 30-1012, 30-1013, 30-1014 and 30-6011.

DateNov 10, 2009 (initiated Sep 24, 2009)
StatusTerminated
ReasonTHERE WERE INCONSISTENCIES NOTICED ON THE INSULATION OVERLAP. THE ELECTRODE HAS A LARGE DIAMETER SHAFT THAT TRANSITIONS TO A SMALLER DIAMETER SHAFT. THE 1ST LAYER OF INSULATION WHICH COVERS THE SMALL DIAMETER SHAFT OVERLAPS WITH THE 2ND LAYER OF INSULATION THAT COVERS THE LARGER DIAMETER SHAFT. ON SOME ELECTRODES IT APPEARS THAT THE OVERLAP IS TOO SHORT TO COVER THE TRANSITION WHICH MAKES THIS
Root causeProcess control
ActionOlsen Medical issued an "Urgent: Device Recall" letter to Consignees beginning September 24, 2009 . Customers were instructed to inventory their stock and for all remaining product immediately discontinue use and return. Consignees were also instructed to contact their customers and notify them of the recall. For further information, contact Olsen Medical at 1-502-772-4272.
Quantity3,617 pieces
DistributionNationwide Distribution -- AZ, FL, IA, MD, MO, NM, OR, NV, TN, TX and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973234MIDAS TOUCH ELECTROSURGICAL ELECTRODE
Iti Medical Technologies, Inc., Oct 27, 1997