Recall Z-1316-2010 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

ARCHITECT System Software v5.00 and v5.10, List Number 5F48-22 and 5F48-23; Part Number 7-204717-01 and 7-204717-02 when used with PM Algorithm LLS Brd, part number 7-94255-01 and PM LLS kit, U10 Chip and Ejectors, part number 7-200788-01. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests utilizing photometry and potentiometric technology, and immunoassay tests utilizi

DateApr 9, 2010 (initiated Sep 11, 2009)
StatusTerminated
ReasonARCHITECT System Software v5.00 and v5.10 are incompatible with ARCHITECT LLS Board. Software patch re-establishes correct parameters.
Root causeSoftware design
ActionTechnical Service Bulletins dated September 11, 2009 were issued to affected consignees. The correction was issued via the Technical Service Bulletins (TSB) to provide a software patch to re-establish the correct parameters on the LLS board. Direct questions to Abbott Laboratories, Inc. by calling 1-972-518-6000.
Quantity11 in the USA; 104 worldwide
DistributionWorldwide.