Recall Z-1156-2010 ยท LPB Cardiac Ablation Percutaneous Catheter

Boston Scientific Corp

CircuCool Fluid Circulation Pump used in the Chilli Cooled Ablation System, Model Number M00480050, manufactured by Barnant ThermoFisher Scientific, Barrington, IL. Distributed by Boston Scientific Corporation, San Jose, CA. The Chilli Cooled Ablation System is indicated for cardiac electrophysiological mapping, delivering diagnostic pacing stimuli and radiofrequency ablation of mappable ventric

DateMar 17, 2010 (initiated Sep 22, 2009)
StatusTerminated
ReasonProduct failure-- Product may cease operation as a result of potential speed control board failure when the device is shut off, and may not turn back on. A component on the circuit board fails unexpectedly because of a power surge, causing a diode to burn out.
Root causeNonconforming Material/Component
ActionThe firm has issued notification letters via Federal express with reply verification tracking forms. A follow up letter and telephone survey as needed will be sent to follow up with non-responders. The firm is asking its consignees to return any recalled product. If there are any questions, please call 1.408.935.4952.
Quantity75 units
DistributionUS and to Spain, France, Germany, Italy, Russia, Poland, the Netherlands, Israel and Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.