ClearanceBrief
Recall Z-0525-2010 ยท KWP Appliance, Fixation, Spinal Interlaminal

Zimmer Inc.

Abbott Spine Thoracolumbar Common, SpdLnk 5.5 Trans Cntr Med 43-51 mm, non-sterile, Manufactured by Abbott Spine, Inc., Austin, TX; Ref. 1302-55M. Note: This lot was not distributed in the United States. The component is used in the stabilization of constructs in the BacFix and InCompass spinal fixation systems. It is used within a temporary construct that assists normal healing and is not inte

DateDec 14, 2009 (initiated Sep 21, 2009)
StatusTerminated
ReasonThe titanium stock used to manufacturer the product is reported to exhibit areas of alloy segregation, which could result in a weakening of the strength of the metal.
Root causeNonconforming Material/Component
ActionConsignees were notified by letter dated September 21, 2009 and instructed to return the affected product to Zimmer, Inc.
Quantity13
DistributionWorldwide Distribution -- Canada, Mexico and South Africa. (Product not distributed in the United States)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002082SPEEDLINK TRANSVERSE CONNECTOR
Spinal Concepts, Inc., Oct 6, 2000