ClearanceBrief
Recall Z-0068-2010 ยท MAF Stent, Coronary

Boston Scientific Corporation

Boston Scientific Liberte Monorail and Over-the-Wire Coronary Stent Systems to be relabeled as VeriFLEX". The Liberte Coronary Stent System consists of a balloon expandable stainless steel stent, pre-mounted on either a Liberte Monorail (MR) or Liberte Over-the-Wire (OTW) delivery system. The Liberte system comes in stent diameters of 2.25, 2.50, 2.75, 3.00, 3.50, 4.00, 4.50 and 5.00 mm and st

DateOct 21, 2009 (initiated Sep 11, 2009)
StatusTerminated
ReasonBoston Scientific initiated a field correction for the Liberte Bare-Metal coronary stent products. They have received reports from cardiac cath labs that TAXUS Liberte Paclitaxel-Eluting (TAXUS Libert Drug-Eluting) coronary stents have been inadvertently selected when the interventional cardiologist intended to implant a Liberte Bare-Metal stent, and Libert Bare-Metal stents have been ina
Root causeLabeling design
ActionBoston Scientific issued an "Urgent Medical Device Field Correction" letter dated September 18, 2009. The letter was addressed to Risk Manger/Field Action Contact. The letter described the product, problem and actions which included renaming the Liberte Bare-Metal stent as VeriFlex" Bare-Metal stent. Enclosed with the letter included the Direction for Field Correction and a Reply Verification Tracking Form. For further information, contact you local Boston Scientific Sales Representative or call 1-800-811-3211.
Quantity391,769
DistributionNationwide Distribution -- United States including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.