ClearanceBrief
Recall Z-0508-2010 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Stryker Howmedica Osteonics Corp.

Secur-Fit MAX Secure-FIT Plus Max Surgical Protocol. Surgical protocols instruct surgeons on proper surgical technique.

DateJan 11, 2010 (initiated Aug 31, 2009)
StatusTerminated
ReasonCertain surgical protocols do not provide specific instructions for initiating power to the calcar planer prior to engagement with bone.
Root causeError in labeling
Action"Urgent Product Correction - Surgical Protocol" letters and Product Accountability Forms were sent to consignees via Fed Ex on August 31, 2009 and September 1, 2009 with return receipt. Notification letters described the issue, protocols affected, potential hazard and harm and risk mitigation. Direct questions about the recall by calling 1-201-972-2100 or Customer Service 1 -866-OR-ASSIST (1 -866-672-7747).
Quantity3,510
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022555ACCOLADE C FEMORAL STEMS
Howmedica Osteonics Corp., Aug 26, 2002