Recall Z-0504-2010 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Stryker Howmedica Osteonics Corp.

Accolade System Surgical Technique; Accolade C - Cemented Hip Stem; Accolade TMZF Cementless Hip System; Accolade HFx Cementless Hip System. Stryker Orthopaedics. Surgical protocols instruct surgeons on proper surgical technique.

DateJan 11, 2010 (initiated Aug 31, 2009)
StatusTerminated
ReasonCertain surgical protocols do not provide specific instructions for initiating power to the calcar planer prior to engagement with bone.
Root causeError in labeling
Action"Urgent Product Correction - Surgical Protocol" letters and Product Accountability Forms were sent to consignees via Fed Ex on August 31, 2009 and September 1, 2009 with return receipt. Notification letters described the issue, protocols affected, potential hazard and harm and risk mitigation. Direct questions about the recall by calling 1-201-972-2100 or Customer Service 1 -866-OR-ASSIST (1 -866-672-7747).
Quantity5,670
DistributionWorldwide Distribution.