Recall Z-0166-2010 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Koodinat M/MP Patient Table used with Angiostar Model 5294009. The device is used as an Angiographic X-Ray System.

DateNov 10, 2009 (initiated Jul 1, 2009)
StatusTerminated
ReasonThe table swivel locking mechanism may not completely engage adn could result in uncontrolled patient movement during transfer resulting in a patient fall.
Root causeLabeling False and Misleading
ActionThe recalling firm issued a Customer Safety Advisory Notice dated July 1, 2009 to affected customers via Update Instructions AX025/09/S to inform customer of the potential issues and provide instructions to prevent occurrence. The firm also provided an addendum to User Manual. Direct questions about the recall to Siemens Medical Solutions USA, Inc. by calling 1-610-219-6300.
Quantity107 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K884912ANGIOSTAR
Siemens Medical Solutions USA, Inc., Feb 16, 1989