Recall Z-0121-2010 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens ARTISTE syngo RT Therapist, Therapist Assist, Therapist Express Basic, and Therapist Express Assist Systems, manufactured by Siemens Medical Solutions, Concord, CA

DateNov 5, 2009 (initiated Sep 10, 2009)
StatusTerminated
ReasonA failed data transfer is possible that may result in data not being recorded in the interfaced system. Siemens Medical Solutions, Inc. USA became aware of this potential issue with its ARTISTE Linear Accelerator's Syngo RT Therapist v4.1 Workspace (containing the Siemens brand of Verify and Record application called PRIMEVIEW) and the LANTIS Oncology Information System product.
Root causeSoftware design
ActionThe firm issued a TH015/09/S Safety Modification syngo RT Therapist 4.1 Customer Advisory letter, dated September 10. 2009. the letter stated the reason for modification, the affected systems, working time, deadline, reimbursement, and reporting. The firm is preparing a software upgrade for the affected products to correct the problem. The letter was sent via certified mail.
Quantity97 units
DistributionWorldwide Distribution -- USA, Spain, Germany, Poland, Australia, Malaysia, New Zealand, the UK, India, Ireland, Egypt, Italy, Belgium, Republic of Korea, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072485ARTISTE MV SA
Siemens Medical Solutions USA, Inc., Dec 27, 2007