Recall Z-1631-2010 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific CRM Corp

Boston Scientific, LATITUDE Communicator, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112. Models 6476 (USA) and 6468 (OUS)

DateMay 20, 2010 (initiated Sep 10, 2009)
StatusTerminated
ReasonBoston Scientific has determined that alert conditions for a limited subset of out-of-range Daily Measurements related to leads may not be recognized by Model 6476 (United States) or Model 6468 (Europe) LATITUDE in-home communicators. If a communicator fails to recognize a valid alert situation, notification for this or subsequent alerts for the same test will not be displayed for physician review
Root causeSoftware design
ActionConsignees were sent on 9/10/09 a Boston Scientific "Urgent Medical Device Information" letter dated September 9, 2009. The letter was addressed to "Dear Doctor". The letter provided the information related to the LATITUDE Patient Management system recall as to Background, Description, Rate of Occurrence, Patient management, Devices Affected and Further Information. Patients were also sent a Boston Scientific letter dated September 9, 2009. The letter addressed Background information, Why are we communicating, What You Should Do and Questions.
Quantity25,656 (USA), 390 (OUS)
DistributionAll States in the US including Puerto Rico United Kingdom, Switzerland, Netherlands, Italy, Germany, France, Finland, Denmark, Belgium, Austria
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.