ClearanceBrief
Recall Z-0146-2010 ยท DXG Computer, Diagnostic, Pre-Programmed, Single-Function

Hospira Inc

Power cord for Q2Plus/Continuous Cardiac Output (CCO)/SO2 Computer, list #s 56711-04-01, 56711-04-03, 5671-04-05, 56711-04-51, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.

DateNov 9, 2009 (initiated Aug 11, 2009)
StatusTerminated
ReasonFire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.
Root causeComponent design/selection
ActionHospira initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and a letter to consignees was sent via UPS. The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.
Quantity204,999 power cords distributed for all devices
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932414CONTINUOUS CARDIAC OUTPUT SYSTEM
Abbott Laboratories, Nov 14, 1994